An Official publication of The Asian Congress of Neurological Surgeons (AsianCNS)

Search Article
Home About us Editorial board Search Ahead of print Current issue Archives Submit article Instructions Advertise Subscribe Contacts Login  Facebook Tweeter
  Users Online: 191 Home Print this page Email this page Small font sizeDefault font sizeIncrease font size  
ORIGINAL ARTICLE
Year : 2018  |  Volume : 13  |  Issue : 4  |  Page : 1066-1070

Cochlear implant device failure in the postoperative period: An institutional analysis


1 Department of Neurosurgery, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, Uttar Pradesh, India
2 Department of ENT and Head and Neck Surgery, BJ Medical College, Ahmedabad, Gujarat, India

Correspondence Address:
Dr. Simple Ravikumar Bhadania
MRA 85, Sanjay Gandhi Post Graduate Institute Campus, Sanjay Gandhi Post Graduate Institute, Rai-Bareilly Highway, Lucknow - 226 014, Uttar Pradesh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ajns.AJNS_93_17

Rights and Permissions

Introduction: As the cochlear implant (CI) surgeries are on rise, there is a compelling need to understand its long-term complications and revision surgery protocols. Our experience in the management of CI devices failure is shared in this paper. Aim: To review the experience in patients who underwent CI device manipulation/explanation in terms of failure rate, etiology, surgical considerations, and preoperative and postoperative auditory and speech outcome. Study Design: This was a retrospective study. Materials and Methods: A retrospective study of 250 patients (201 children and 49 adults) with normal cochlea at a tertiary care center from June 2004 to June 2014 was done. All cases were implanted multichannel devices via Veria technique of CI surgery. Preoperative assessment, surgical considerations, and postoperative auditory and speech outcomes were analyzed. Preoperative and postoperative auditory/speech outcomes were analyzed using Category of Auditory Perception (CAP) and Speech Intelligibility Rating (SIR) scores. Results: Reimplantation rate was 4%. The causes of revision CI surgery were hard device failure (n = 3), surgical site infection (n = 3), magnet displacement (n = 2), and electrode extrusion (n = 2). In one patient, recurrent cutaneous infection on the implanted site ultimately resulted in reimplantation in the opposite ear after multiple surgical interventions on the same side. The preoperative and postoperative CAP and SIR scores showed improvement in the postoperative period with P < 0.05 as compared with the paired t- test. Conclusions: Preoperative counseling for device failure should always be emphasized. The success rate is high in revision surgery with good performance in the postoperative audiological outcome. There is a compelling need for an agreed international definition of CI failure and the adoption of uniform reporting protocols.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed235    
    Printed12    
    Emailed0    
    PDF Downloaded34    
    Comments [Add]    