Local balloon-assisted navigation of a microcatheter into an aneurysm during intracranial aneurysmal coiling: A dunk shot technique


1 Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi, Japan
2 Department of Neurosurgery, Aichi Medical University, Nagakute, Aichi, Japan
3 Neuroendovascular Therapy Center; Department of Neurosurgery, Aichi Medical University, Nagakute, Aichi, Japan

Date of Submission19-Jan-2021
Date of Decision19-Feb-2021
Date of Acceptance17-Mar-2021
Date of Web Publication28-May-2021

Correspondence Address:
Dr. Tomotaka Ohshima
Neuroendovascular Therapy Center, Aichi Medical University, 1-1 Yazakokarimatoa, Nagakute 480-1195, Aichi
Japan
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ajns.AJNS_26_21

Rights and Permissions
  Abstract 

Objectives: Various endovascular approaches have been reported for patients with intracranial aneurysms. However, the safety of navigating a microcatheter into the aneurysm remains debatable. In this study, we evaluated a novel method “dunk shot technique,” in which a proximal balloon and a local balloon are used for navigation of a microcatheter into an aneurysm under challenging situations. Materials and Methods: We have reported two cases of unruptured internal carotid artery–superior hypothalamic artery aneurysm. An 8-F balloon-attached catheter was used as the guiding catheter. A local balloon catheter for neck remodeling and a microcatheter for coil insertion were navigated around the aneurysmal neck region. When it seemed difficult to navigate a microcatheter into an aneurysm, both the guiding balloon and a local balloon catheter were used for assistance. After inflation of the guiding balloon, the local balloon was inflated partially to negotiate the tip of the microcatheter. Results: The uncontrollable tip of the microcatheter could be intentionally moved by the local balloon. Conclusions: We evaluated the effectiveness of the balloon-assisted technique for the navigation of a microcatheter in cases with challenging anatomy. A little bit of direct effect to the tip of the microcatheter by a local balloon could produce effective outcomes under the proximal flow arrest.

Keywords: Balloon-assisted technique, catheterization, coil embolization, intracranial aneurysm


How to cite this article:
Ohshima T, Kawaguchi R, Matsuo N, Miyachi S. Local balloon-assisted navigation of a microcatheter into an aneurysm during intracranial aneurysmal coiling: A dunk shot technique. Asian J Neurosurg 2021;16:384-6

How to cite this URL:
Ohshima T, Kawaguchi R, Matsuo N, Miyachi S. Local balloon-assisted navigation of a microcatheter into an aneurysm during intracranial aneurysmal coiling: A dunk shot technique. Asian J Neurosurg [serial online] 2021 [cited 2021 Jun 25];16:384-6. Available from: https://www.asianjns.org/text.asp?2021/16/2/384/316879


  Introduction Top


Navigating a microcatheter into the aneurysm is necessary for successful coil embolization; however, it is often difficult because of the underlying anatomical condition. Rough manipulation may introduce a premature intraprocedural rupture that may prove fatal. We encountered two challenging anatomical conditions: one had a ledge effect due to distal aneurysmal neck, while another had a curved parent artery. Because we advanced the 8-F balloon-attached guiding catheter (BGC) for the treatment of all anterior circulation aneurysms, a local balloon could be inflated at the precise desired position under the proximal flow arrest by the BGC. Here, we have reported a novel technique that employed a proximal balloon and a local balloon for the navigation of a microcatheter into an aneurysm under challenging anatomical situations.


  Case Report Top


Case 1

A previously healthy 62-year-old female was diagnosed with an unruptured right internal carotid artery (ICA)–superior hypothalamic artery aneurysm [Figure 1]a. The patient received an endovascular coil embolization under general anesthesia. An 8-F Optimo BGC (Tokai Medical Products, Aichi, Japan) was advanced into her right ICA via the right femoral artery. First, the Transform SC 4 × 7 Balloon Catheter (Stryker, Kalamazoo, MI, USA) was navigated around the aneurysmal neck region using the Chikai 14 Microguidewire (Asahi Intecc, Aichi, Japan). Next, the Phenom 17 Microcatheter (Medtronic, Irvine, CA, USA) was navigated into the aneurysm using the Traxess Microguidewire (Terumo, Tokyo, Japan). However, the tip of the Phenom 17 was prevented from navigation due to the ledge effect at the distal aneurysmal neck region [Figure 1]b. Therefore, we induced the local balloon-assisted technique at this point. The proximal balloon of the BGC was inflated before the local balloon inflation. The local balloon was partially inflated from the distal side of the aneurysmal neck and then withdrawn [Figure 1]c. The tip of the microcatheter could be successfully navigated into the aneurysm [Figure 1]d. A schematic drawing of the technique is illustrated in [Figure 2].
Figure 1: (a) An angiography of the working angle * Aneurysm (b) the tip of the microcatheter (arrow) could not exceed the distal aneurysmal neck portion due to the ledge effect between the microguidewire and the microcatheter. (c) After the balloon inflation (arrowhead), the tip of the microcatheter (arrow) could exceed the gap. (d) Finally, the tip of the microcatheter (arrow) could be navigated into the aneurysm

Click here to view
Figure 2: Schematic illustration of the local balloon-assisted technique in Case 1. (a) The arrow indicates the ledge effect between the wire and the catheter at the distal aneurysmal neck portion. (b) The expanded balloon allows the tip of the catheter to exceed the corner

Click here to view


Case 2

A 54-year-old female was diagnosed with an unruptured left ICA–superior hypothalamic artery aneurysm [Figure 3]a. The patient received an endovascular coil embolization under general anesthesia. An 8-F Optimo BGC (Tokai Medical Products) was advanced into her left ICA via the right femoral artery. Next, the Scepter XC 4 × 11 Balloon Catheter (Terumo) was navigated around the aneurysmal neck with the Chikai 14 Microguidewire (Asahi Intecc). Then, the Phenom 17 Microcatheter (Medtronic), which was S-shaped, was navigated into the aneurysm using the Traxess Microguidewire (Terumo). However, the tip of the Phenom 17 faced the opposite direction toward the neck along with the curved parent artery. It could not be turned by controlling the microguidewire [Figure 3]b. Therefore, we induced the local balloon-assisted technique. The proximal balloon of the BGC was inflated before inflating the local balloon, and the local balloon was partially inflated at the aneurysmal neck position [Figure 3]c. As a result, the curvature of the microcatheter overturned and the tip became positioned facing the aneurysmal neck. Finally, the tip of the microcatheter could be navigated into the aneurysm [Figure 3]d. A schematic drawing of this technique is illustrated in [Figure 4].
Figure 3: (a) An angiography of the working angle * Aneurysm (b) the tip of the microcatheter (arrow) could not face toward the aneurysm. (c) After the balloon inflation (arrowhead), the tip of the microcatheter could face toward the aneurysm. (d) Finally, the tip of the microcatheter (arrow) could be navigated into the aneurysm

Click here to view
Figure 4: Schematic illustration of the local balloon-assisted technique in Case 2. (a) The bent microcatheter faced the opposite side of the aneurysm and fixed along with the curvature of the parent artery (arrow). (b) The expanded balloon allows the tip of the catheter to face the aneurysm

Click here to view



  Discussion Top


Some past studies used the 8-F BGC and found it useful for coil embolizations.[1],[2],[3],[4] The main reason for using the BGC was for insurance in case of an intraprocedural rupture of the aneurysm. Meanwhile, a blood flow control during coil deployment into the aneurysm was another benefit of using the 4th adjunctive technique, followed by balloon-assistance, stent-assistance, and a double-catheter technique.[3] We believe that the temporary proximal flow arrest by the 8-F BGC functions to prevent excessive flow into the aneurysm when the local balloon is inflated at the distal portion from the aneurysmal neck region.

The uses of both the proximal and distal balloons during coil embolization have been reported. Ansari et al.[4] reported that the proximal flow control was effective for subsequent accurate deployment of the local balloon. They reported a case of repeatedly ruptured, fragile, small ICA aneurysm. The aneurysm was successfully embolized using the double-balloon-trapping technique. In the local balloon-assisted technique used in the present case, there was no need for a full expansion of the distal balloon in comparison with that in the double-balloon-trapping technique.

Parasellar ICA aneurysms are often challenging to treat because the complicated shape of the microcatheter tip is occasionally required for the tortuous parent artery. In our country, a flow diverter is allowed to use only for a broad-necked aneurysm. A variety of microcatheter shaping methods including those under three-dimensional (3D) angiographic guidance, with the use of a 3D printer, and in vivo microcatheter printing method have been recently reported.[5],[6],[7] However, even after the precise shaping is achieved, navigating the microcatheter remains challenging. Therefore, we considered that a little bit of direct touch with a microcatheter at the local site could be effective to assist its navigation into the aneurysm, such as by dunk shot of the basketball. We thought that this dunk shot technique was quite useful for the cases which required making a sharp-angled bent of the tip of the microcatheter [Figure 5].
Figure 5: (a) An angiography of the working angle of the Case 1. Yellow lines indicate the ideal bent of the tip of the microcatheter. The subtended angle is 60 degrees. (b) An angiography of the working angle of the Case 2. Yellow lines indicate the ideal bent of the tip of the microcatheter. The subtended angle is 40°

Click here to view


However, the local balloon-assisted technique offers some disadvantages, and this study has some limitations. First, this was a retrospective analysis conducted at a single center with only two cases. Second, the risk of thromboembolic events was greater with this technique than that with the conventional one. Stent-type devices such as Comaneci temporary embolization assist device (Rapid Medical, Yokneam, Israel) are not available in our country. Finally, this technique requires a more extensive caliber arterial sheath than the conventional technique. Thus, careful and secure hemostasis of the access site is mandatory in this situation.


  Conclusions Top


We have reported the use of balloon-assisted technique for the navigation of a microcatheter in cases of challenging anatomy. A little bit of direct touch with a microcatheter at the local site was found to be effective in assisting its navigation into the aneurysm, such as a dunk shot of the basketball. The proximal flow control by BGC is important to advance an inflated local balloon to avoid excessive flow into the aneurysm when the local balloon is placed distal to the aneurysmal neck region.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Tajima H, Ohshima T, Goto S, Yamamoto T, Shimato S, Nishizawa T, et al. Usefulness of the 8Fr Optimo as a guiding catheter during endovascular treatment for intracranial aneurysms: Original article. J Neuroendovasc Ther 2015;9:16-21.  Back to cited text no. 1
    
2.
Ohshima T, Dash C, Belayev A, Yamamoto T, Goto S, Kato Y. 8-F balloon guide catheter for embolization of anterior circulation aneurysms: An institutional experience in 152 patients. Nagoya J Med Sci 2017;79:435-41.  Back to cited text no. 2
    
3.
Ohshima T, Miyachi S, Matsuo N, Kawaguchi R, Niwa A, Maejima R, et al. Efficacy of the proximal balloon flow control method for endovascular coil embolisation as a novel adjunctive technique: A retrospective analysis. Interv Neuroradiol 2018;24:375-8.  Back to cited text no. 3
    
4.
Ansari A, Ohshima T, Goto S, Yamamoto T, Ishikawa K, Kato Y. Double-balloon trapping for coil embolization of ruptured internal carotid artery aneurysm: A novel technique. Asian J Neurosurg 2019;14:873-7.  Back to cited text no. 4
[PUBMED]  [Full text]  
5.
Toyota S, Fujimoto Y, Iwamoto F, Wakayama A, Yoshimine T. Technique for shaping microcatheter tips in coil embolization of paraclinoid aneurysms using full-scale volume rendering images of 3D rotational angiography. Minim Invasive Neurosurg 2009;52:201-3.  Back to cited text no. 5
    
6.
Namba K, Higaki A, Kaneko N, Mashiko T, Nemoto S, Watanabe E. Microcatheter shaping for intracranial aneurysm coiling using the 3-dimensional printing rapid prototyping technology: Preliminary result in the first 10 consecutive cases. World Neurosurg 2015;84:178-86.  Back to cited text no. 6
    
7.
Ohshima T, Imai T, Goto S, Yamamoto T, Nishizawa T, Shimato S, et al. A novel technique of microcatheter shaping with cerebral aneurysmal coil embolization: In vivo printing method. J Neuroendovasc Ther 2017;11:48-52.  Back to cited text no. 7
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]



 

Top
 
 
  Search
 
<
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)  

 
  In this article
   Abstract
  Introduction
  Case Report
  Discussion
  Conclusions
   References
   Article Figures

 Article Access Statistics
    Viewed96    
    Printed0    
    Emailed0    
    PDF Downloaded20    
    Comments [Add]    

Recommend this journal